AQ1000 / AQ2000
ALTERNATING PRESSURE
BUBBLE PAD AND PUMP SYSTEM
USER MANUAL
Important: Do not operate the Mattress System
without first reading and understanding this manual!
Save this manual for future use.
The latest version of this manual can be found online
AQ1000-INS-LAB-RevD13
INTRODUCTION
Use this manual for setup and operation of the Lumex
AQ1000 / AQ2000 Alternating Pressure Pad and Pump
System. Read all instructions before using the pad and
pump system. Save this manual for future reference.
INTENDED USE OF THIS DEVICE
The intended use of the Alternating Pressure Pad
and Pump System, in either a homecare or long-term
care setting, is the treatment of pressure ulcers while
optimizing user comfort, and pain management as
prescribed by a physician.
Contraindication
WARNING: DO NOT use this product in the presence of
flammable anesthesia. There is a possible fire hazard
when this product is used with certain oxygen delivery
equipment.
1
Use nasal cannula, face mask, or /2 length oxygen tent
to deliver oxygen in the presence of this product.
DO NOT use a full length oxygen tent that extends past
the top surface of the mattress with this product.
IMPORTANT SAFETY PRECAUTIONS
The safety statements presented in this chapter refer
to the basic safety information that should be observed
by those using the Alternating Pressure Pad and Pump
System. There are additional safety statements in other
chapters or sections, which may be the same as or similar
to the following, or specific to the operations.
DANGER: Indicates an imminent hazard situation that, if
not avoided, will result in death or serious injury.
WARNING: Indicates a potential hazard situation or
unsafe practice that, if not avoided, could result in
death or serious injury.
AQ1000-INS-LAB-RevD13
3
CAUTION: Indicates a potential hazard or unsafe
practice that, if not avoided, could result in minor or
moderate personal injury.
s NOTICE: Indicates a potential hazard or unsafe practice
that, if not avoided, could result in product or property
damage.
Info: Provides application recommendations or other useful
information to ensure that you get the most from your
product.
DANGER
To reduce the risk of electrocution:
DANGER: Always unplug this product immediately after
using.
DANGER: Do not use this product while bathing.
DANGER: Do not place or store this product where it
can fall or be pulled into a tub or sink.
DANGER: Do not place this product in or drop into
water or other liquid.
DANGER: Do not reach for a product that has fallen into
water. Unplug it immediately.
WARNING
To reduce the risk of burns, electrocution, fire,
or personal injury:
WARNING: Do not leave this product unattended when
plugged in.
WARNING: Always use close supervision when this
product is used by, on, or near children or those who
require close supervision.
WARNING: Use this product only as intended and
described in this manual. Do not use attachments or
accessories not recommended by GF Health Products,
Inc..
4
AQ1000-INS-LAB-RevD13
WARNING: Never operate this product if
a) it has a damaged cord or plug.
b) it is not working properly.
c) it has been dropped or damaged.
d) it has been dropped into water.
Return the product to your Graham-Field authorized
distributor.
WARNING: Keep the cord away from heated surfaces.
WARNING: Never block the air openings of this product
or place it on a soft surface, such as a bed or couch,
where the openings may be blocked. Keep the air
opening free of lint, hair, and other similar particles.
WARNING: Never drop or insert any object into any
opening or hose.
WARNING: Connect this product only to a properly
grounded power outlet.
WARNING: Maximum patient weight capacity for this
product is 275 lb (125 kg), EVENLY DISTRIBUTED.
WARNING: Notice for California Customers- California
Proposition 65 WARNING: This product contains a
chemical known to the State of California to cause
cancer and reproductive or developmental harm.
NOTICE
s NOTICE: The pump can be used only with the
accompanying bubble pad. Do not use it for any other
purpose.
AQ1000-INS-LAB-RevD13
5
EQUIPMENT SYMBOLS
Attention! Consult accompanying documents (this manual)
“BF”symbol: Indicates that this product is in accordance with the
degree of protection against electric shock for type BF equipment
Grounding terminal
Class II (for UL only)
Disposal of Electrical & Electronic Equipment (WEEE):
Do not treat this product as household waste. For more detailed
information with regard to returning and recycling this product,
please consult your local city office, household waste disposal
service, or Graham-Field authorized distributor.
The system has been tested and successfully approved with the following
standards:
U.S.
UL 1431
Europe
EN 60601-1, EN 60601-1-2, EN 55011 Class B,
EN 61000-3-2, EN 61000-3-3, IEC 801-2, IEC 801-3,
IEC 801-4, IEC 801-5
6
AQ1000-INS-LAB-RevD13
PRODUCT DESCRIPTION
BUBBLE PAD AND PUMP SYSTEM
air tubes
bubble pad
pump
bubble pad and pump system
PUMP
front panel
air tube connectors
wire
hangers
fuses
(220V only)
pump (AQ2000 shown)
AQ1000-INS-LAB-RevD13
7
FRONT PANEL
AQ1000
ON/OFF switch
air tube connectors
AQ1000 pump front panel
AQ2000
ON/OFF
switch
pressure
control dial
AQ2000 pump front panel
Front panel pressure control dial (AQ2000 only): The
pressure control dial controls the air pressure output. To
increase pressure, turn the dial clockwise. To decrease
pressure, turn the dial counter-clockwise.
8
AQ1000-INS-LAB-RevD13
INSTALLATION
UNPACKING
1. Before unpacking, check for obvious damage to the
carton or its contents. If damage is evident, please
notify the carrier and your Graham-Field authorized
distributor.
2. Remove all loose packing from the carton.
3. Carefully remove all components from the carton.
4. Inspect all components. If damage is evident, please
notify the carrier and your Graham-Field authorized
distributor.
PUMP AND BUBBLE PAD INSTALLATION
1. Remove all sheets, blankets, and pillows from the bed.
2. Place the bubble pad on the bed mattress with the
tubing at the patient's feet.
wire hanger
foot end
bubble pad
mattress
pump
(AQ2000 shown)
installation
3. Fold open the wire hangers on the back of the pump as
shown above.
4. Use the hangers to hang the pump on the bed's foot end,
as shown above; adjust the hangers to place the pump
in the most upright position (or place the pump on a flat
surface easily accessible to the caregiver and/or doctor).
s NOTICE: Place the device in a position where the
caregiver and/or doctor can access it easily.
AQ1000-INS-LAB-RevD13
9
5. Connect one end of the air tubes to the bubble pad:
slide the ends of the air tubes over the connectors on
the top of the foot end of the bubble pad.
6. Connect the other end of the air tubes to the pump
connectors; ensure that the ends of the air tubes are
firmly fitted over the bubble pad and pump connectors
before continuing.
s NOTICE: Ensure that the air tubes are not kinked or
tucked under the bubble pad or mattress.
WARNING: Locate all cords so that they will not be
stepped on, tripped over, or otherwise subjected to
damage or stress.
WARNING: Do not use a household extension cord if
the electrical cord does not reach the power outlet. Use
of an improper extension cord could result in fire and
electric shock. If an extension cord must be used, use
a three conductor cord with ground, properly wired,
in good electrical condition, and keep it as short as
possible.
WARNING: This device—which MUST be grounded—is
equipped with a cord that has a grounding wire with
a grounding plug. The plug must be plugged into an
outlet that is properly installed and grounded.
WARNING: Ensure that the local power voltage is
appropriate for the pump unit.
7. Plug the power cord into a properly grounded electrical
outlet.
10
AQ1000-INS-LAB-RevD13
OPERATION
1. Turn the ON/OFF switch to the ON position. The pump
will begin to deliver air into the bubble pad.
2. Replace all sheets, blankets, and pillows on the bed.
Info: if the pressure level seems low, check for any leakage
in the air tubes or bubble pad. Replace any damaged tubes. If
necessary, contact your Graham-Field authorized distributor
for repair or replacement.
AQ2000 ONLY: PRESSURE ADJUSTMENT
Adjust the pressure control dial for comfort.
Info: For maximum effectiveness, patient should sink into
bubble pad to reduce pressure points. Ensure that patient
does not bottom out.
QUICK DISCONNECT
Pull the power plug from the wall connector to disconnect
the device quickly.
CPR FUNCTION
If CPR must be performed on the patient while the bubble
pad is in use, to immediately deflate the bubble pad, pull
the air tubes off the bubble pad or pump connectors.
AQ1000-INS-LAB-RevD13
11
MAINTENANCE
CLEANING
Perform the following cleaning procedures before use and
as necessary thereafter.
Pump
Wipe the pump with a clean, damp cloth and mild
detergent, and keep it away from dust. Air dry.
s NOTICE: Do not use phenolic products or corrosive or
powdered cleansers to clean the pump.
s NOTICE: Do not immerse or soak the pump.
Bubble pad
Wipe the bubble pad with a clean, damp cloth and mild
detergent. The bubble pad may also be cleaned using a
10% solution of sodium hypochlorite (bleach) diluted in
water. Air dry all parts thoroughly before use.
s NOTICE: Do not use phenolic products to clean the
bubble pad.
s NOTICE: After cleaning, air dry the bubble pad without
direct exposure to sunlight.
GENERAL MAINTENANCE
1. Check main power cord for abrasion or excessive wear.
2. Check bubble pad for signs of wear or damage. Ensure
air tubes are firmly secured to bubble pad and pump
connectors.
3. Check the air tubes for kinks or breaks. For
replacement, please contact your Graham-Field
authorized distributor.
12
AQ1000-INS-LAB-RevD13
Fuse replacement (220V system only)
tool needed: small screwdriver
1. If you suspect a blown fuse, disconnect the plug from
the wall outlet immediately.
fuse
fuse cover
fuse location
2. Use a small screwdriver to remove the cover of the fuse
holder (fuse locations are shown in picture above).
3. Insert a new fuse of the correct rating (T1A/250V).
4. Replace the fuse holder cover. Ensure that fuse holder
cover is securely installed.
STORAGE
1. Fold the bubble pad loosely.
2. Store bubble pad and pump in a dust-free environment
with no exposure to direct sunlight.
AQ1000-INS-LAB-RevD13
13
SPECIFICATIONS
Model
AQ1000 (Standard)
AQ2000 (Deluxe)
120V: 120V AC, 60 Hz, 1A
Power Supply
220V: 230VAC, 50 Hz, 7W
120V: N/A
Fuse Rating
220V: T1A, 250V
1.9 psi (13.10 kPa)
1.0 - 2.0 psi (6.89-13.79 kPa)
Pressure
120V: 4.5 LPM
220V: 5.0 LPM
120V: 4-5 LPM
220V: 5-6 LPM
Air Output
120V: 5 min.
Cycle Time
220V: 6 min.
10 ft. (304.8 cm)
Power Cord
Length
8.5 x 4.25 x 3.62 in.
(21.59 x 10.80 x 9.19 cm)
9.75 x 5.12 x 3.88 in.
(24.77 x 13.00 x 9.86 cm)
Pump Unit Size
2.9 lb. (6.38 kg)
3.5 lb. (7.7 kg)
Pump Unit Weight
78 x 34 x 2.5 in.
(198.12 x 86.36 x 6.35 cm)
76 x 34 x 2.5 in.
(193.04 x 86.36 x 6.35 cm)
Bubble pad Size,
Inflated
275 lb. (125 kg), EVENLY DISTRIBUTED
Weight Capacity
14 GA anti-bacterial, flame-retardant vinyl
Bubble pad
Material
2 each: 5 ft. (152.4 cm) long clear PVC
Tubing
Contact your Graham-Field dealer or visit Graham Field's
Replacement
Parts
Info: Specifications are subject to change without notice.
14
AQ1000-INS-LAB-RevD13
WARRANTY
GF Health Products, Inc. warrants the Lumex AQ1000 /
AQ2000 Alternating Pressure Pad and Pump System for
defects in workmanship and materials for the terms listed
below.
Pump: one year
Pad:
90 days
If a product is deemed to be under warranty, GF Health
Products, Inc. shall provide, at its option, (1) replacement
of any defective part or product or (2) a credit of the
original selling price made to GF Health Products, Inc.’s
initial customer. The warranty does not include any labor
charges incurred in replacement part(s) installation or
any associated freight or shipping charges to GF Health
Products, Inc.
The warranties contained herein contain all the
representations and warranties with respect to the
subject matter of this document, and supersede all
prior negotiations, agreements and understandings with
respect thereto. The recipient of this document hereby
acknowledges and represents that it has not relied on any
representation, assertion, guarantee, warranty, collateral
contract or other assurance, except those set out in this
document.
AQ1000-INS-LAB-RevD13
15
GF Health Products, Inc.
2935 Northeast Parkway
Atlanta, Georgia 30360
tel: 770-368-4700
fax: 770-368-2386
© 2008 GF Health Products, Inc.
|